Dental Curing Light with Optional Tooth Whitening,Lesion Detection and Soft Tissue Management (STM) 3W Diode Laser AttachmentsInstructions for Use
10Setup and Operation Instructions Compatibility of Existing Sapphire® and Sapphire Plus Lights and Devices The labeling on the curing handpiece and S
11Light Meter (Radiometer) Use the Sapphire Plus light meter to validate the power density (light intensity or mW/cm2) output of the curing handpiece
12Curing Mode (continued)To change from the C1 to the C2 timer, press the LEFT and RIGHT arrow keys on the CONTROL WHEEL. The display windows show “C2
13Bleach (Whitening) ModeAfter the system power is switched ON ( I ), use the RIGHT arrow key to advance to the BLEACH mode. Or, if in the CURE mode,
14Whitening InstructionsLUMIBRITE™ Chairside Whitening System contains stabilized hydrogen peroxide for maximum whitening ability and fluoride for over
1513. Press the Start or pistol activator button to start the bleaching process. The number 60 (minutes) appears in the Program window and sec
16STM (Soft Tissue Management) 3W Diode Laser Mode Settings NOTE: DO NOT ATTEMPT TO OPERATE THE SAPPHIRE® PLUS STM 3W DIODE LASER SYSTEM WITHOUT FIRST
17STM 3W Diode Laser Mode (continued)Use the CW/P button to toggle between the CW (CONTINUOUS WAVE) AND PULSE modes.Press the READY button in either m
18STM 3W Diode Laser Mode (continued)The STM 3W Diode Laser unit may be deactivated at any time by pressing the Emergency “STOP” button on the STM 3W
19Service Minor problems can be quickly identified and resolved. These are usually related to the electrical power connection or light guide connectio
2Safety Precautions (Read first before proceeding)WarningBefore connecting and operating the Sapphire® Plus, read all safety instructions. When used pr
20Lamp Replacement Instructions WARNING: ONLY A QUALIFIED SERVICE TECHNICIAN MAY REPLACE THE SYSTEM LAMP. PROTECTIVE EYEWEAR MUST BE WORN. NEVER POINT
21Fuse Replacement WARNING: ONLY A QUALIFIED SERVICE TECHNICIAN MAY REPLACE SYSTEM FUSES. 1. If the Sapphire® Plus source fails to operate after pow
22The Den-Mat® Holdings, LLC Sapphire STM 3W Diode Laser SpecificationSTM Diode Laser Unit:Style: Plug-In accessory for Sapphire® Plus SystemL
23System Warranty When used under normal operating conditions as described in this manual, the Sapphire® Plus system (main system, light guide, and c
©2012 Den-Mat® Holdings, LLC. World Rights Reserved. 823057810 03/12COCALL 1-800-445-0345 TO ORDERCALL YOUR DEN-MAT® HOLDINGS, LLC TECHNICAL SALES REP
3Important Start-up Instructions (Follow these instructions prior to using the system)1. Attach the AC Power Cord to the power inlet on the Sapphire
4Table of ContentsSafety Precautions ...
5General Description The Sapphire® Plus is a multi-purpose platform that combines the capabilities of a high-intensity xenon arc lamp with specially-d
6Basic System ContentsSapphire® Light Source Operator’s ManualLight Guide and Tip AC Power CordCuring Light Protective Glasses (2 pair) Com
791187155A5B5C326410Controls and Functions 1. Sapphire Plus Light Source 2. Sapphire Plus Control Panel: Includes control wheel and select button, mo
8Control PanelDisplay GlossaryDisplay Meaning Next StepNo HP No handpiece Install handpieceCh HP Choose handpiece Press RIGHT arrow to change mo
9Accessories EyewearThe protective eyewear provided with the Sapphire® Plus must be worn by the operator, assistants, and patient when the system is